Abstract

We present a new chemiluminescence detector, with solution channels that have been machined into a Teflon disk and sealed with a sapphire window. The configuration of the flow cell can be conveniently modified by replacing the Teflon disk. A comparison of some existing and novel designs, using the chemiluminescence reaction of morphine with acidic potassium permanganate and the bioluminescence reaction of ATP with the commercially available "BacTiter-Glo" reagent, has revealed that a serpentine channel allows greater quantities of light to be captured than a spiral channel, due to more efficient mixing of the analyte and reagent solutions within the cell.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.