Abstract
To detect NO in samples of the headspace of static cores, a sample-mixing unit was developed which made it possible to detect NO in small-volume samples (50μl to 10 ml) by a continuous flow chemiluminescence NO analyzer. The sample-mixing unit enables the detector of the NO analyzer to reach a steady-state response. The detection limit was 3.5 pmol NO in a sample, which is equivalent to 8.5 ppbv NO in a 10-ml sample or 85 ppbv No in a 1-ml sample at room temperature. The coefficient of variation between replicates varied between 6.4% at low NO levels and 0.6% at high NO levels. The linear response was good, theR2 value was above 0.99 even in the lowest detection range. The sample-mixing unit in combination with an NO analyzer provides a sensitive, reliable, easy to use method for the detection of NO in incubation systems with small headspace volumes such as static cores.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.