Abstract

Improved sensitivity of the cold-vapour atomic absorption method for mercury can be obtained by equilibrating the reduced sample with a small volume of air at 90°C. An automated system has been developed that has a detection limit of 1 ng Hg l -1. By changing the reducing conditions three species of mercury can be differentiated and determined, inorganic mercury, arylmercury compounds such as phenylmercury(II) chloride, and alkylmercury compounds such as methylmercury(II) chloride. Speciation of mercury in natural waters is possible.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.