Abstract

Aim: Quantification of stereoisomers in biological matrices is of pivotal importance for drug development. Supercritical fluid chromatography paired with chiral stationary phases is the gold standard for resolution of enantiomers. However, this technique often proves inadequate for resolution of polar stereoisomers. Materials & methods: A combination of achiral chemical derivatization with supercritical fluid chromatography using chiral stationary columns to improve enantiomeric resolution is described. Results: Separation of four possible stereoisomers of linerixibat was achieved after derivatization with 3N HCl in n-butanol within 12min (case1). Derivatization with acetic, propionic, butyric, isobutyric, valericand isovaleric anhydrides significantly improved the separation of stereoisomers (case 2 and3) within 10min. The best stereoisomeric resolution was achieved using valeric and isovaleric anhydrides.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call