Abstract
Despite the potential benefits of photopolymerization-based additive manufacturing, photochemical reactions in free-radical polymerization rarely proceed to completion, leading to the accumulation of residual monomer in polymer networks. In this novel study, we examined an array of methacrylate polymers for denture bases (>60 % Bis-EMA and 15–25 % proprietary methacrylic oligomer), orthodontic appliances (>70 % proprietary methacrylic oligomer, <20 % glycol methacrylate and <5 % pentamethyl-piperidyl sebucate) and implant surgical guides (≥75 % Bis-EMA and 30–50 % diurethane dimethacrylate, mixture of isomers) using headspace gas chromatography-mass spectrometry. Experimental data reported herein show that a substantial proportion of chemical compounds observed in the methacrylates decreased with post-illumination and the applied ethanol treatment. In the absence of residual methyl methacrylate, other potentially toxic acrylic esters were observed thus emphasizing the need to thoroughly scrutinize additively manufactured dental devices prior to their use. In the long term, standards for medical devices in dentistry could be revised to reflect the current trends in biomaterials and precursors they are generated from.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.