Abstract

In the RA-1000, a random-access discrete analyzer, an inert fluorocarbon fluid is used to prevent interaction and carryover. Production-model instruments were evaluated in two laboratories with respect to determination of glucose, creatinine, total protein, inorganic phosphorus, cholesterol, alkaline phosphatase, lactate dehydrogenase, creatine kinase, gamma-glutamyltransferase, and aspartate and alanine aminotransferases. Within-run, among-run, and day-to-day (for 15 days) precision was assessed, and results were correlated with those obtained by the methods routinely in use in our departments. Precision was excellent, correlation acceptable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.