Abstract
A commercially available porous silica monolithic column (Onyx Monolithic Si, 100mm×4.6mm I.D.) was ‘in-column’ covalently functionalised with 2-hydroxyethyliminodiacetic acid (HEIDA) groups, and applied to the simultaneous and rapid separation of alkaline earth and transition metal ions, using high-performance chelation ion chromatography (HPCIC). With a 0.3mM dipicolinic acid (DPA) containing eluent, the baseline separation of various common transition and heavy metal ions and the four alkaline earth metal ions could be achieved in under 14min with a flow rate of just 0.8mL/min. Detection was achieved using spectrophotometric detection at 540nm after post-column reaction (PCR) with 4-(2-pyridylazo)-resorcinol (PAR). Significant effects from variation of eluent nature, concentration and temperature upon selectivity and retention were demonstrated with the new monolithic silica chelating phase. Under optimised conditions (0.165M LiNO3 eluent, pH 2.5), peak efficiencies of 54,000, 60,000 and 64,000N/m, for Zn2+, Mn2+ and Cd2+, respectively, were recorded, far exceeding that previously reported for IDA based chelation ion exchange columns.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.