Abstract
A new method for the determination of lanthanides based on chelation and ion chromatography with absorbance detection is described. γ-Aminobutyrohydroxamate resin or its derivative, N-methyl- γ-aminobutyrohydroxamate resin in a column was used to concentrate and separate lanthanides from the alkali and alkaline earth metals and other matrix components. Various complexing agents were investigated as possible eluants. The influence of a sample matrix and parameters important for quantitative analysis are discussed. Using γ-aminobutyrohydroxamate resin as concentrator, and potassium nitrate (0.02 M) as matrix eliminator, with the mobile phase of oxalic acid (0.03 M, pH 4.5)-diglycolic acid (0.005 M), and 4(2-pyridylazo) resorcinol (6 × 10 −5 M)- ZnEDTA(1 ×10 −4 M) as postcolumn reagent, most of the lanthanides can be separated and determined in the complex matrix samples. The detection limit was found to be 2.5 ng ml −1 for these elements by concentrating 25 ml of sea water.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.