A new PVC membrane electrode for the triiodide ion based on a charge-transfer complex of iodine with 7,16-dibenzyl-1,4,10,13-tetraoxa-7,16-diazacyclooctadecane as a membrane carrier was prepared. The electrode exhibits a Nernstian response for triiodide ions over a wide concentration range (1.0 x 10(-1)-1.0 x 10(-5) M) with a slope of 59.3 +/- 0.9 mV decade(-1) and a detection limit of 6.3 x 10(-6) M. It has a response time of 30 s and can be used for at least 3 months without any divergence in the potential. The potentiometric response is independent of the pH, in the pH range 1.6 - 10.0. The proposed electrode has shown very high selectivity for the triiodide ion over a wide variety of other anions. This electrode was successfully applied as an indicator electrode in the potentiometric titration of ascorbic acid and hydroquinone from pharmaceutical preparations as well as ascorbic acid in orange juice and dissolved O2 in tap water.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.