Abstract

Active matrix flat panel dosimeters (AMFPD's) have a high density of charge trapping centers (/spl sim/3/spl times/10/sup 15//cm/sup 3/) which lead to both lag and ghosting in reconstructed dose measurements. Lag is due to charge release within the detector and is a function of the lifetime of the trapping centers within the detector. Ghosting, defined as excess charge not produced by charge release, is a phenomenon believed to be caused by reduced charge trapping probability due to the trapped charge population density being greater after exposure to a radiation field than previous to the exposure. This work examines the characteristics of a large AMFPD (26 cm /spl times/ 26 cm) detector for long exposure times (up to 1 minute) and high doses (up to 3 Gy). This detector was operated in a direct detection mode. Nonlinear response functions were mapped as a function of both exposure time and dose at a beam energy of 6 MV. Additionally, the impulse response functions (IRF's) were also measured at 6 MV for the same set of exposure times and doses used for the measurement of the nonlinear mapping. These IRF's are shown to be functions of both exposure time and dose delivered to the detector. The reproducibility of the dose measurements is discussed.