Abstract

The beam current in an external-beam PIXE–PIGE is difficult to accurately measure due to ionization along the beam path in the atmosphere. Charge integration was measured using a homemade Faraday cup, and assessed by the peak area of Ar Kα X-rays, which were induced by protons near the sample. The X-ray peak integral from a thin Fe reference sample, which was positioned between the exit window and the Faraday cup, was determined to evaluate the performance of the homemade Faraday cup. Moreover, the effects of different membrane filters and samples with different elements on the beam current measurement of the aforementioned methods were studied by placing different blank films or reference standards behind a reference Mn target. The results indicated that the charge measurement of the homemade Faraday cup was reliable for external PIXE–PIGE analysis of aerosol samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.