Abstract

Tight requirements on the Large Synoptic Survey Telescope point spread function (PSF) demand sensor contribution to PSF be both small and well characterized. The sensor PSF is determined by the lateral charge diffusion on the drift path from the photon conversion point to the gates. The maximum drift path occurs for photons converted at the window, for blue optical photons in particular. Charges generated at the window surface undergo "worst case" charge spreading and the blue optical PSF is used to characterize the sensor's PSF. Different techniques for charge diffusion characterization have been developed, each with its own systematics and measurement difficulties. A new way to measure charge diffusion using an X-ray source is presented. We demonstrate the effectiveness and limitations of our technique and discuss relation of charge diffusion value obtained with X-ray measurements to sensor PSF.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.