Abstract

Validation of the various NRC blackbody sources below 962 °C has been hampered by the lack of a radiation thermometer with sufficient sensitivity and adequately small target size. Recently, progress was made in this area by having access to a NIST RT1550, a radiation thermometer operating at 1.55 μm with a thermoelectrically cooled InGaAs detector, a 3 mm nominal target size, a measuring range of 150 °C to 1064 °C, and a noise-equivalent temperature of 0.5 mK at 420 °C. Having the RT1550 for several months allowed sufficient time to intercompare our various blackbodies, both fixed-point and variable-temperature. It is these intralaboratory comparison results that are reported here.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.