Specifying and reducing the uncertainty is very important to all kinds of measurements. In order to have a better understanding of the noise mechanism responsible for lowering the quality of mass/force measurements, the Allan variation method is applied to investigate the noise performance of a commercial ultra-microbalance installed on different noise reduction platforms. It turns out that the marble table provides a better noise isolation environment for mass measurements than an optical table. The optimal integration time is found to be 100 sec ~ 200 sec, with a lowest deviation of 0.07 μg. A different data treatment simulating the ABA load/unload cycle is also applied with or without a delay time for signal integration. The consistency between Allan deviation and the ABA simulation plots points out that the optimal integration time is applicable either in single or cyclic mass measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.