Abstract

IntroductionCranioplasty is performed by placing an artificial plate in place of a patient's native skull bones to repair post-craniectomy defects after trauma. Implanted materials can range from titanium to synthetic polyether derivatives and are produced by multiple manufacturers. There are few studies characterizing complications associated with these cranioplasty plates to date. We aimed to quantify and categorize complications of these devices using a national federal database.MethodsThe Manufacturer and User Facility Device Experience (MAUDE) database was queried for all entries reported under the category "plate, cranioplasty, preformed, non-alterable" with the additional product code GXN between the time range from September 1, 2010, to September 1, 2020. After data extraction, each of the entries was screened for duplicates and tabulated into different categories of complications. Additionally, product information such as the plate manufacturer was extracted from each entry.ResultsThe search yielded 329 unique event reports. The most frequent complications were infection (39%), followed by incorrectly fitting implants (30%) and implant breaks (6%). Other major complications included cerebrospinal fluid leakage and cerebral edema (5%), wound dehiscence (5%), and migration of hardware (3%). The brands associated with the most entries in the database were the Synthes (DePuy Synthes Companies, Massachusetts, United States) polyetheretherketone (PEEK) patient-specific implants (PSI) (57%), the Biomet (Zimmer Biome, Indiana, United States) hard tissue replacement-polyetherketoneketone (HTR-PEKK) patient-matched implant (PMI) (23%), and the AccuShape PEEK PSI (MedCAD, Dallas, USA) (5%).ConclusionsInfection and improperly fitting implants appear to be the two most frequent complications of cranioplasty plates. The goals of future research should include the prevention of plate infections as well as improving techniques to custom-fit implantable devices.

Highlights

  • Cranioplasty is performed by placing an artificial plate in place of a patient's native skull bones to repair post-craniectomy defects after trauma

  • Other major complications included cerebrospinal fluid leakage and cerebral edema (5%), wound dehiscence (5%), and migration of hardware (3%)

  • The brands associated with the most entries in the database were the Synthes (DePuy Synthes Companies, Massachusetts, United States) polyetheretherketone (PEEK) patient-specific implants (PSI) (57%), the Biomet (Zimmer Biome, Indiana, United States) hard tissue replacementpolyetherketoneketone (HTR-PEKK) patient-matched implant (PMI) (23%), and the AccuShape PEEK PSI (MedCAD, Dallas, USA) (5%)

Read more

Summary

Objectives

We aimed to quantify and categorize complications of these devices using a national federal database

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.