Abstract

Zero-thickness interface models are developed to describe the encapsulation of microbubble contrast agents. Two different rheological models of the interface, Newtonian (viscous) and viscoelastic, with rheological parameters such as surface tension, surface dilatational viscosity, and surface dilatational elasticity are presented to characterize the encapsulation. The models are applied to characterize a widely used microbubble based ultrasound contrast agent. Attenuation of ultrasound passing through a solution of contrast agent is measured. The model parameters for the contrast agent are determined by matching the linearized model dynamics with measured attenuation data. The models are investigated for its ability to match with other experiments. Specifically, model predictions are compared with scattered fundamental and subharmonic responses. Experiments and model prediction results are discussed along with those obtained using an existing model [Church, J. Acoust. Soc. Am. 97, 1510 (1995) and Hoff et al., J. Acoust. Soc. Am. 107, 2272 (2000)] of contrast agents.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.