Abstract
AbstractCharacterization of the STA1600 CCD for the high‐resolution spectrograph Potsdam Echelle Polarimetric and Spectroscopic Instrument (PEPSI) is presented. We tested the fixed‐pattern noise reduction efficiency with sufficient number of master flat fields and show that the noise reduction follows the expected value. The lateral charge diffusion is measured with photon transfer technique versus CCD binning factor and by measuring the change of spectral lines width versus their amplitude. The charge diffusion effect versus wavelength follows the behavior of the back‐illuminated CCD while also showing a strong positional dependence on the CCD.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.