Abstract

A precise model of a Si(Li) detector ORTEC® model SLP-06165P-OPT-0.5 was created for beta spectra calculations using the Monte Carlo (MC) code MCNPX™. Detector parameters were determined from X-ray radiograms obtained with a film and a Timepix detector. The MC model of the detector was validated by comparison of calculated and experimental full-energy peak efficiencies in the energy range from 5 to 136 keV using a range of point-like photon emitting radionuclide standards. A comparison of measured and calculated beta spectrum of a radionuclide Pm-147 is presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.