Abstract
The achievable spatial resolution and the contrast transfer function (CTF) are key parameters characterizing an X-ray microscope. We measured the spatial resolution and the contrast transfer function of the transmission X-ray microscope (TXM) at the electron storage ring BESSY II. The TXM uses the radiation of an undulator source and operates under partially coherent illumination conditions. For spatial resolutions down to 25 nm, our measurements of the CTF's are in good agreement with theoretical CTF data for partial coherence. With higher resolution zone plate objectives, we measured a spatial resolution (half-pitch) of 11 nm in 1st and 3rd order of diffraction. However, with these objectives the stray light level increases significantly.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.