Abstract

The purpose of this project was to quantify reciprocity law failure (RLF) for mammography screen-film systems. Three widely used screen-film systems were evaluated: the Kodak MinR 2000 system. Fuji UM Mammo Fine screen and Fuji UM MA HC film, and Agfa MR Detail screen and Agfa Mammoray MR5 film. The logit algorithm that linearizes logistic curve shapes was utilized to characterize film sensitometric response. Different values of mammographic phantom thickness, tube current, and kVp were used to vary screen-film exposure rates. RLF was quantified by examining the dependence of logit parameters (maximum and minimum film density, curve shift, and slope) on exposure rate. The shift of the logit curve was found to be a good indicator of the screen-film system speed, while the slope of the logit curve is affected by the RLF. RLF leads to changes in film contrast as well as speed. For the range of exposure rates measured (50-fold), screen-film contrast and speed varied by factors of 2 and 3.5, respectively. Film contrast decreased as exposure rate increased. The greatest changes were observed with the Kodak MinR 2000 screen-film system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.