Abstract
A new device delivering intrapulmonary percussive ventilation (IPV), called Impulsator® (Percussionaire Corporation, Sandpoint, ID, USA), has recently been introduced in an effort to provide effective clearance and to promote homogeneity of ventilation in the lungs of patients with cystic fibrosis. In order to optimize the treatment based on its use, a better understanding of its functioning is still necessary. In fact, up to now, a complete characterization of this device has not been carried out, thus reducing its effective utilization in clinical practice. With the aim of overcoming this lack, in this study, data concerning flow and pressure delivered during in vitro IPV were acquired under different combinations of device settings and respiratory loads. Quantitative information was obtained about the physical variables administered by the device like percussive frequency, ratio of inspiratory to expiratory time, flow and pressure magnitudes and volume exchanged. The analysis of the data determined the relations among these variables and between them and the mechanical loads, laying the basis for an optimal clinical application of the device.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.