Abstract

In order to evaluate an atomic clock, it is important to determine the main frequency shifts caused by external fields, device imperfections, etc. Scanning the frequency of the main oscillator, the Ramsey fringe is obtained and used to determine the main shifts. The gravitational shift (10-17), second order Doppler (1.65×10-13), Black-Body radiation shift (2.9×10-14), quadratic Zeeman effect (5×10-13), Rabi Pulling (1.3×10-13) and cavity pulling (1.3×10-13) have been evaluated. The short term stability, (6.6±0,2)×10-13, was obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.