Abstract
The micropallet array system uses a pulsed laser to release pallets tens of microns to hundreds of micrometers in size from a larger array, enabling selective isolation of single cells adherent to the pallets. We characterize the laser-based release of pallets with respect to pallet array and laser parameters. The threshold laser energy required for pallet release increases linearly with the area of the pallet in contact with the underlying glass substrate. The spacing of the pallets within an array as well as the thickness or height of the pallet does not impact the energy required to release a pallet. Delivery of multiple laser pulses decreases the energy/pulse required for pallet release when the pallets were 100 microm or greater on a side. In addition to the square pallets, complex structures such as cantilevers and spirals could be released without damage using the pulsed laser. Identification of the pallet-array variables influencing the energy required for pallet release as well as strategies to minimize this energy will prove critical in optimizing the release of pallets with cells on the arrays.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.