Abstract

The effect of experimental conditions such as gas flow rates, detector and column temperatures on sulphur response are discussed. It is concluded that the response of the flame photometric detector is affected by the O/H ratio and the total gas flow. Column temperature has an important influence upon S2 emission. It is shown that the sulphur response of the FPD is dependent on the compound containing the sulphur.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.