Abstract

This work evaluates the temporal stability, spatial stability, batch reproducibility and dose rate dependence of a new cosolvent-free polymer gel dosimeter optimized for use with x-ray computed tomography readout. Temporal and spatial stability investigations reveal the new gel formulation should be imaged between 15-36 hours after irradiation. Intra- and inter-batch reproducibility were found to be excellent over the entire range of doses examined. A dose rate dependence was found for gels irradiated with machine dose rates between 100-600MU/min. An intensity-modulated radiation therapy treatment validation is also presented to illustrate an example clinical application using the new gel formulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.