Abstract

Standard Reference Material (SRM) 2940 is a cuvette-shaped, Mn-ion-doped glass, recommended for use for relative spectral correction of emission and day-to-day performance validation of steady-state fluorescence spectrometers. Properties of this standard that influence its effective use or contribute to the uncertainty in its certified emission spectrum were explored here. These properties include its photostability, absorbance, dissolution rate in water, anisotropy, temperature coefficient of fluorescence intensity, and fluorescence lifetimes. Long and short lifetime components of the fluorescence displayed different emission spectra, making the certified spectrum useful with fluorescence instruments employing continuous excitation only. The expanded uncertainties in the certified spectrum are about 5% around the peak maximum at 620 nm, using an excitation wavelength of 412 nm. The SRM also exhibits a strong resistance to photodegradation, with no measurable decrease in fluorescence intensity even after 17 h of irradiation with the visible light from a Xe lamp.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.