Abstract

To prevent the particle generation in a process and to reduce the number of killer size particles, it is obviously important to get the information of exact particle size. The most popular tool for the process particle monitor is a laser surface scanner, whose detection sensitivity is defined by calibration using a PSL standard particles. We will report the difference between the real size of particles on process monitor wafers an the measured size by the laser surface scanner using a PSL calibration curve. For example, the real particle size is 14 micrometers but a machine A displays 164 micrometers and a machine B displays 4 micrometers on an AlCu-deposited wafer. The machine A displayed size is smaller than our measured size, in the range 0.1 to 10 micrometers . On the other side, beyond 10 micrometers size is displayed larger. To identify the source of particles, and to reduce and keep the number of killer size particles below the upper control limit, a laser surface scanner should display more accurate particle size which can be seen on the deposition layer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.