Abstract

The Large Hadron Collider (LHC) at CERN will use high field superconducting magnets operating in pressurized superfluid helium (He II) at 1.9 K. Cold safety valves, with their inlet in direct contact with the He II bath, will be required to protect the cold masses in case of a magnet resistive transition. In addition to the safety function, the valves must limit their conduction heat load to the He II to below 0.3 W and limit their mass leakage when closed to below 0.01 g/s at 1.9 K with 100 mbar differential pressure. The valves must also have a high tolerance to contaminating particles in the liquid helium. The compliance with the specified performance is of crucial importance for the LHC cryogenic operation. An extensive test program is therefore being carried out on prototype industrial valves produced by four different manufacturers. The behavior of these valves has been investigated at room temperature and at 77 K. Precise heat load and mass leak measurements have been performed on a dedicated test facility at superfluid helium temperature. Results of cold and warm tests performed on as-delivered valves are presented.KeywordsLarge Hadron ColliderHeat LoadSafety ValveValve InletSuperfluid HeliumThese keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.