Abstract
For accurate and simultaneous imaging of fast neutrons (FNs) and prompt gamma rays (PGs) produced during proton therapy, the selection of a highly performant detector material is crucial. In this work, a promising candidate material known as organic glass scintillator (OGS) is characterized for this task. To this end, a precisely-timed source of neutrons and Bremsstrahlung radiation produced by the nELBE facility was used to study the light output and neutron/gamma ray pulse shape discrimination (PSD) properties of a 1 × 1 × 20 cm3 OGS bar with double-sided readout. Furthermore, the energy, timing, and depth-of-interaction (DOI) resolutions of 1 × 1 × 10 cm3 and 1 × 1 × 20 cm3 OGS and EJ-200 bars were characterized with radioactive sources. For electron-equivalent energies above 0.5 MeVee, OGS was found to have excellent PSD capabilities (figure-of-merit above 1.27), energy resolution (below 12%), coincident time resolution (below 500 ps), and DOI resolution (below 10 mm). This work establishes the data analysis methods required for hybrid FN/PG imaging using OGS, and demonstrates the materials' excellent performance for this application.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.