Abstract
The paper deals with the characterization of the neutron output during operation the neutron generator in the pulsing mode. The characterization is performed by a cadmium covered He-3 detector. First, the adequacy of a cadmium covered He-3 detector to study the pulsed mode operation of neutron generator is assessed using Monte Carlo simulations in MCNP6 code. Second, the pulsing mode operation of the D-D P-385 Thermo Fisher Scientific neutron generator with a maximal neutron output of 7 x 106 n/s is studied. The study shows the dependence of neutron emission and pulse shapes on the frequency, duty factor, high voltage and beam current.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.