Abstract

Hollow-fiber hemodialyzers containing cellulose-based membranes have been shown to produce positive results with the Limulus amoebocyte lysate test. This study was undertaken to determine whether endotoxin was causing the reaction. Rinses from 45 parallel-plate and hollow-fiber dialyzers from eight different manufacturers were tested before and after treatment with cellulase, using three lysates and four Limulus amoebocyte lysate methods. In addition, four in vitro cellular methods--human leukocytic pyrogen, lymphocytic activating factor, peritoneal macrophage, and arginase release--were used to evaluate endotoxin activity. The substance causing the reaction was identified by chromatographic methods. Results indicate that the Limulus amoebocyte lysate reactive material is cellulose derived and not pyrogenic.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.