Abstract

Two coaxial and a low-energy HPGe detector were characterized with Monte Carlo simulations, using the geant4 toolkit. The geometry of the detectors, including the dimensions of the crystal and the internal structural parts, were initially taken from the factory specifications and from X-ray radiographies, and were later fine-tuned. The detector response functions, with special emphasis on the absolute full-energy peak efficiencies and peak-to-total ratios, were calculated and compared to experimental data taken at different measurement geometries. Between 150 keV and 11 MeV an agreement within 1–2 standard deviation has been achieved, whereas systematic deviations were experienced at lower energies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.