Abstract

We report the characterization of a balanced ultrafast photodetector. For this purpose, we use a recently developed time-domain laser-based vector network analyzer (VNA) to determine the common-mode rejection ratio (CMRR) of the device under test. This includes the frequency-domain response above the single-mode frequency of the coaxial connector. Although the balanced photodetector has a nominal bandwidth of 43 GHz, it generates voltage pulses with frequency components up to 180 GHz. We obtain a CMRR of better than 30 dB up to 70 GHz and better than 20 dB up to 110 GHz. The laser-based measurements are compared with the measurements using a digital sampling oscilloscope and with the frequency-domain measurements using a conventional VNA. We obtain good agreement between the three techniques with the laser-based method providing the largest measurement bandwidth, although it also constitutes the most complicated characterization setup.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.