Abstract

A method for the determination of the molecular weight distribution of heparins using heparin oligosaccharides as calibration standards and high-performance size exclusion liquid chromatography has been developed. An organic microparticulate stationary phase (TSK 3000), an aqueous mobile phase and UV detection at 206 nm were adopted. Commercial heparins, low-molecular-weight heparins and polysulphated mucopolysaccharides with heparin-like activities were characterized by this procedure. Average molecular weights [weight average ( M w ), number average ( M n ), z average ( M z )] and molecular weight distribution (cumulative weight fraction and differential weight fraction) were calculated from the chromatographic data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.