Abstract

Novel traceable analytical methods and reference gas standards were developed for the detection of trace-level ammonia in biogas and biomethane. This work focused on an ammonia amount fraction at an upper limit level of 10 mg m-3 (corresponding to approximately 14 μmol mol-1) specified in EN 16723-1:2016. The application of spectroscopic analytical methods, such as Fourier transform infrared spectroscopy, cavity ring-down spectroscopy, and optical feedback cavity-enhanced absorption spectroscopy, was investigated. These techniques all exhibited the necessary ammonia sensitivity at the required 14 μmol mol-1 amount fraction. A 29-month stability study of reference gas mixtures of 10 μmol mol-1 ammonia in methane and synthetic biogas is also reported.