Abstract

Sofosbuvir is a direct acting antiviral medication used to treat Hepatitis C viral infection. The present study focuses on the degradation behavior of the drug under various stress conditions (hydrolysis, oxidative, thermal and photolytic) as per International Conference on Harmonization (ICH Q1A (R2)) guidelines. A high performance liquid chromatographic system (HPLC) was used to develop a selective, precise and accurate method for separating all the degradation products. The separation was achieved on a Sunfire™ C18 (150mm×4.6mm×5μm) stationary phase with a mobile phase of 10mM ammonium acetate (pH 5.0) buffer and acetonitrile in gradient elution mode. A quadrupole-time of flight mass analyzer equipped with an electrospray ionization technique was used to propose the structural information based on the MS/MS and accurate mass measurements. Seven degradation products were identified and characterised by LC-ESI-QTOF–MS/MS. In silico toxicity of the drug and its degradation products was determined using TOPKAT and DEREK toxicity prediction softwares. The proposed method was validated as per the ICH Q2 guidelines.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call