Abstract

Objective. With the introduction of Cherenkov imaging technology on the Halcyon O-ring linear accelerator platform, we seek to demonstrate the imaging feasibility and optimize camera placement. Approach. Imaging parameters were probed by acquiring triggering data Cherenkov image frames for simplistic beams on the Halcyon and comparing the analyzed metrics with those from the TrueBeam platform. Camera position was analyzed by performing 3D rendering of patient treatment plans for various sites and iterating over camera positions to assess treatment area visibility. Main results. Commercial Cherenkov imaging systems are compatible with the pulse timing of the Halcyon, and this platform design favorably impacts signal to noise in Cherenkov image frames. Additionally, ideal camera placement is treatment site dependent and is always within a biconical zone of visibility centered on the isocenter. Visibility data is provided for four treatment sites, with suggestions for camera placement based on room dimensions. Median visibility values were highest for right breast plans, with values of 80.33% and 68.49% for the front and rear views respectively. Head and neck plans presented with the lowest values at 26.44% and 38.18% respectively. Significance. This work presents the first formal camera positional analysis for Cherenkov imaging on any platform and serves as a template for performing similar work for other irradiation platforms. Additionally, this study confirms the Cherenkov imaging parameters do not need to be changed for optimal imaging on the Halcyon. Lastly, the presented methodology provides a framework which could be further expanded to other optical imaging systems which rely on line of sight visibility to the patient.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call