Abstract

A 4.7 ×4.7×9.5 mm 3 CdZnTe Frisch collar device was characterized through probing the device with a highly collimated Cs 137 662 keV gamma ray source. In a systematic series of experiments, the detector was probed along the length and width with a Cs 137 gamma ray source using a 43.0 mm long Pb-collimator with a 0.6 mm circular hole. The detector was probed along the central line under different operating voltages of 1200, 1000, 800, 600 and 400 V. The experimental results correlated well to charge collection calculations for a modeled device with the same size and operating conditions. It was proved that, unlike the planar configuration, the charge collection efficiency profile along the length of Frisch collar device is considerably improved. The CdZnTe raw materials for this study were acquired from Redlen Technologies, and the Frisch collar device was fabricated and characterized at S.M.A.R.T. Laboratory at Kansas State University.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.