Abstract

This paper presents the measurement of the sub‐micrometer deformation of biocompatible materials induced by mechanical and thermal loading with a newly developed stereoscopic microscope 3D digital image correlation (DIC) system. This technique combines a 3D DIC system based on two cameras in a stereoscopic setup with a stereoscopic microscope. This paper describes measurements of displacements in the range of 100 microns with an error of 0.1%, which is several times better than the accuracy provided by other comparable systems. The examined micro‐sized specimens are cut out of biocompatible material used for the construction of a pneumatic ventricular assist device. The results include a 3D surface profile, analysis of the sample surface deformations, and distribution of surface strains.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.