Abstract

A radiochromic silicone based dosimeter has been fabricated that allows for three-dimensional (3D) dose verification. This paper characterizes the silicone dosimeter and verifies previous results found by other research groups. Using a recipe found to be dose rate independent, we identify the spectra of the dosimeter irradiated to different doses and verify that the response from different dose rate deliveries is constant within uncertainty. The reproducibility of a response from identically irradiated and stored cuvettes is also investigated. A noticeable difference in response indicates that there is an inhomogeneous distribution of active ingredients due to the viscous nature of the dosimeter during fabrication. Irradiating multiple cuvettes to the same dose in different fractions shows that the response decreases as the number of fractions increases.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.