Abstract
The line width (often synonymously used for critical dimension, CD) is a crucial parameter in integrated circuits. To accurately control CD values in manufacturing, a reasonable CD reference material is required to calibrate the corresponding instruments. We develop a new reference material with nominal CDs of 160 nm, 80 nm, and 40 nm. The line features are investigated based on the metrological scanning electron microscope which is developed by the National Institute of Metrology (NIM) in China. Also, we propose a new characterization method for the precise measurement of CD values. After filtering and leveling the intensity profiles, the line features are characterized by the combination model of the Gaussian and Lorentz functions. The left and right edges of CD are automatically extracted with the profile decomposition and k-means algorithm. Then the width of the two edges at the half intensity position is regarded as the standard CD value. Finally, the measurement results are evaluated in terms of the sample, instrument, algorithm, and repeatability. The experiments indicate efficiency of the proposed method which can be easily applied in practice to accurately characterize CDs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.