Abstract

As a highly deployed field instrument for the detection of narcotics, explosives, and chemical warfare agents, drift tube ion mobility spectrometry relies heavily upon the performance of the ionization source and mechanism of ion beam modulation. For this instrumental platform, ion chemistry plays a critical role in the performance of the instrument from a sensitivity and selectivity perspective; however, a range of instrumental components also occupy pivotal roles. Most notably, the mechanism of ion modulation or ion gating is a primary contributor to peak width in a drift tube ion mobility experiment. Unfortunately, physical ion gates rarely perform perfectly, and in addition to serving as physical impediments to ion transmission, their modulation also has undesirable field effects. Using a recently developed modulated, non-radioactive X-ray source, we detail the performance of an ion mobility spectrometry (IMS) system that is free of a gating structure and utilizes the pulsed nature of the modulated X-ray source (MXS) for both ion generation and initiation of the IMS experiment. After investigating the influence of pulse duration and spatial X-ray beam width on the analytical performance of the instrument, the possibility of using multiplexing with a shutterless system is explored. By increasing ion throughput, the observed multiplexing gain compared to a signal-averaged spectrum approaches the theoretical maximum and illustrates the capability of the MXS-IMS system to realize significant signal to noise improvements.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call