Abstract
This article presents the performance test results of the CFVAD3 continuous flow blood pump in an artificial human circulation system. The CFVAD3 utilizes magnetic bearings that support a thin pancake impeller, the shape of which allows for a very compact pump whose total axial length is less than 5 cm with a radial length of about 10 cm. This gives a total volume of about 275 cc. The impeller itself has 4 vanes with a designed operating point of 6 L/min at 100 mm Hg of differential pressure and 2,000 rpm. The advantages of magnetic bearings, such as large clearance spaces and no mechanical wear, are elaborated upon. Furthermore, bearing model parameters such as load capacity and current gains are described. These parameters in conjunction with the operating conditions during testing are then used to estimate the fluid forces, stiffness, and damping properties while pumping. Knowledge of these parameters is desirable because of their effects on pump behavior. In addition, a better plant model will allow more robust control algorithms to be devised that can boost pump performance and reliability.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.