Abstract
In this work we present measurements of the sensitivity of a low noise microwave receiver developed with the aim to detect a feeble quantum electrodynamics effect known as dynamical Casimir effect. We study the performance of the receiver when it is connected to a transmission line ending with an antenna coupled to a resonant microwave cavity. The noise temperature of the receiver is measured with the method of the variable temperature load resistor. The noise generated by the cavity is measured in a similar way. An equivalent input noise of ( 2.0 ± 0.2 ) × 10 - 22 W / Hz is measured at the receiver input. This is equivalent to a sensitivity of approximately 100 photons, well below the expected signal from the quantum vacuum photons.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.