Abstract
A liquid scintillator (LS) based on linear alkyl benzene (LAB) solvent has been characterized using multiple radiation sources. The results confirm that boron-loaded LAB is suitable for neutron detection in a gamma ray environment. To study indirectly the LAB pulse shape discrimination (PSD) capability between neutrons and gamma rays, a dissolved 212Pb source emitting alpha and beta particles was used to emulate the conditions in a mixed radiation field for detecting neutrons in the presence of a high gamma ray background. The quenching factor depends on the alpha energy and increases from 10 to 25 as the alpha energy decreases from 10 to 1MeV. 10B loaded LAB-based LS has been tested in a neutron beam of energy which is equal to 14.56meV. The observed peak at 60keVee is attributed to the absorption of neutrons. Our results show that a boron-loaded LAB-based scintillator is a sensitive medium for neutron detection in a relatively large background of gamma rays. A neutron detector could be achieved with a figure of merit (FOM) of 1.75.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.