Abstract

In this work we present the characterization of a large area $200~\hbox{mm} \times 200~\hbox{mm}$ EJ-444 scintillation detector to be used for monitoring gross alpha and beta activity in tap water plants. Specific tests were performed in order to determine the best setup to read-out the light from the detector side. The possibility to stack many detectors and get a compact device with total active area of the order of $1~\hbox{m}^2$ has been explored. Alpha/beta discrimination, efficiency and homogeneity tests were carried out with alpha and beta sources. Background from ambient radioactivity was measured as well. Alpha/beta real-time monitoring in drinking water is a goal of the EU project TAWARA_RTM.