Abstract

We present a method to accurately control the photon energies for hard X-ray Self-seeding schemes with a single crystal monochromator in transmissive geometry. The energy calibration is performed by measuring which pairs of the machine pitch and yaw angles for different crystallographic planes reflect the X-ray at the same wavelength. The free parameters of an analytical formula for the self-seeding energies are determined by fitting the observed intersections and the normalized derivative with respect to the pitch and yaw angles in the observed intersections. The method requires a hard X-ray spectrometer, but it does not rely on its absolute energy calibration. Instead, identifying the self-seeded energies above the SASE background or the monochromatic notches within the SASE bandwidth is sufficient for the calibration.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call