Abstract

A HPGe detector was modelled and characterized by means of experimental measurements and Monte Carlo simulations. The simulation based on Monte Carlo GEANT4 was used to characterize the potential dead layer/inactive materials and Full Energy Peak Efficiency (FEPE). The calibration of FEPE for the HPGe detector was conducted in the γ-ray energy range from 88 keV to 1332 keV, using eight standard point sources of gamma radiation: 137Cs, 133Ba, 109Cd, 65Zn, 60Co, 57Co, 54Mn, and 22Na. The efficiency dependence curves were obtained by changing the source to detector distance from 8 cm to 12 cm. a remarkable similarity in FEPE for low energy measurements ranging from 88 keV to 122 keV was obtained between the simulation results and the experimental measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.