Abstract
Water cluster ion intensity and distribution is affected by source conditions in direct sample analysis (DSA) ionization. Parameters investigated in this paper include source nozzle diameter, gas flow rate, and source positions relative to the mass spectrometer inlet. Schlieren photography was used to image the gas flow profile exiting the nozzle. Smaller nozzle diameters and higher flow rates produced clusters of the type [H + (H(2)O)(n)](+) with greater n and higher intensity than larger nozzles and lower gas flow rates. At high gas flow rates, the gas flow profile widened compared with the original nozzle diameter. At lower flow rates, the amount of expansion was reduced, which suggests that lowering the flow rate may allow for improvements in sampling spatial resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Society for Mass Spectrometry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.