Abstract

A digital x-ray imaging system was designed for small-animal studies. This system is a fiber-optics taper-coupled imaging system with two CCD arrays uniquely jointed. The x-ray source of the system has a small focal spot of 20 microm. This digital imaging system contains specially designed shelves to provide magnification levels, ranging from 1.5x to 5x. The system is characterized in terms of its properties of spatial resolution. An observer-based spatial resolution measurement was conducted with a line-pair target and a sector test pattern. The modulation transfer function of the system, with different magnifications, was studied by use of a 10-microm lead slit. The average resolutions at 50% and 5% modulations at 1x magnification were measured as 3.9 and 8.4 lp/mm, respectively, where lp indicates line pairs. With 5x magnification, the 50% and the 5% modulations provided 13.2- and 29.9-lp/mm, respectively, average spatial resolutions. The measurements showed consistency between the two individual CCD arrays; the difference in resolution between the two CCDs is less than 1%, even at high magnifications.